PIERL 119 [2024] PIERL 118 [2024] PIERL 117 [2024] PIERL 116 [2024] PIERL 115 [2024] PIERL 114 [2023] PIERL 113 [2023] PIERL 112 [2023] PIERL 111 [2023] PIERL 110 [2023] PIERL 109 [2023] PIERL 108 [2023] PIERL 107 [2022] PIERL 106 [2022] PIERL 105 [2022] PIERL 104 [2022] PIERL 103 [2022] PIERL 102 [2022] PIERL 101 [2021] PIERL 100 [2021] PIERL 99 [2021] PIERL 98 [2021] PIERL 97 [2021] PIERL 96 [2021] PIERL 95 [2021] PIERL 94 [2020] PIERL 93 [2020] PIERL 92 [2020] PIERL 91 [2020] PIERL 90 [2020] PIERL 89 [2020] PIERL 88 [2020] PIERL 87 [2019] PIERL 86 [2019] PIERL 85 [2019] PIERL 84 [2019] PIERL 83 [2019] PIERL 82 [2019] PIERL 81 [2019] PIERL 80 [2018] PIERL 79 [2018] PIERL 78 [2018] PIERL 77 [2018] PIERL 76 [2018] PIERL 75 [2018] PIERL 74 [2018] PIERL 73 [2018] PIERL 72 [2018] PIERL 71 [2017] PIERL 70 [2017] PIERL 69 [2017] PIERL 68 [2017] PIERL 67 [2017] PIERL 66 [2017] PIERL 65 [2017] PIERL 64 [2016] PIERL 63 [2016] PIERL 62 [2016] PIERL 61 [2016] PIERL 60 [2016] PIERL 59 [2016] PIERL 58 [2016] PIERL 57 [2015] PIERL 56 [2015] PIERL 55 [2015] PIERL 54 [2015] PIERL 53 [2015] PIERL 52 [2015] PIERL 51 [2015] PIERL 50 [2014] PIERL 49 [2014] PIERL 48 [2014] PIERL 47 [2014] PIERL 46 [2014] PIERL 45 [2014] PIERL 44 [2014] PIERL 43 [2013] PIERL 42 [2013] PIERL 41 [2013] PIERL 40 [2013] PIERL 39 [2013] PIERL 38 [2013] PIERL 37 [2013] PIERL 36 [2013] PIERL 35 [2012] PIERL 34 [2012] PIERL 33 [2012] PIERL 32 [2012] PIERL 31 [2012] PIERL 30 [2012] PIERL 29 [2012] PIERL 28 [2012] PIERL 27 [2011] PIERL 26 [2011] PIERL 25 [2011] PIERL 24 [2011] PIERL 23 [2011] PIERL 22 [2011] PIERL 21 [2011] PIERL 20 [2011] PIERL 19 [2010] PIERL 18 [2010] PIERL 17 [2010] PIERL 16 [2010] PIERL 15 [2010] PIERL 14 [2010] PIERL 13 [2010] PIERL 12 [2009] PIERL 11 [2009] PIERL 10 [2009] PIERL 9 [2009] PIERL 8 [2009] PIERL 7 [2009] PIERL 6 [2009] PIERL 5 [2008] PIERL 4 [2008] PIERL 3 [2008] PIERL 2 [2008] PIERL 1 [2008]
2012-01-13
A Calibration Procedure for Two-Port VNA with Three Measurement Channels Based on T-Matrix
By
Progress In Electromagnetics Research Letters, Vol. 29, 35-42, 2012
Abstract
A simplified calibration procedure using T-matrix concept is presented for two-port vector network analyzer (VNA) with three measurement channels. Compared with Short-Open-Load-Thru (SOLT) calibration method based on 10-term error model where 10 error terms must be solved and saved at each frequency, the proposed method need define fewer characteristic variables. Moreover, a length-unknown 50Ω line can be used instead of a random single-port standard, for example, substituting Load standard with 50Ω line in SOLT calibration procedure. Via the simplified calibration procedure, the scattering parameters of a two-port device under test (DUT) can be finally obtained. Experimental verification is carried out, and good agreement is observed.
Citation
Wei Zhao, Hong-Bo Qin, and Li Qiang, "A Calibration Procedure for Two-Port VNA with Three Measurement Channels Based on T-Matrix," Progress In Electromagnetics Research Letters, Vol. 29, 35-42, 2012.
doi:10.2528/PIERL11112102
References

1. Rumiantsev, A. and N. Ridler, "VNA calibration," IEEE Microwave Magazine, Vol. 9, No. 3, 86-99, 2008.
doi:10.1109/MMM.2008.919925

2. Rytting, D., "Network analyzer error models and calibration methods," RF & Microwave Measurement for Wireless Application (ARFTG/NIST Short Course Notes), 1996.

3. Ferrero, A., V. Teppati, M. Garelli, and A. Neri, "A novel calibration algorithm for a special class of multiport vector network analyzers," IEEE Trans. on Microw. Theory and Tech., Vol. 56, No. 3, 693-699, 2008.
doi:10.1109/TMTT.2008.916952

4. Marks , R. B., "Formulations of the basic vector network analyzer error model including switch-terms," Proc. 50th ARFTG Microwave Measurements Conf. Fall, 115-126, 1997.

5. Vandenberghe , S., D. Schreurs, G. Carchon, B. Nauwelaers, and W. De Raedt, "Identifying error-box parameters from the twelve-term vector network analyzer error model," Proc. 60th ARFTG Microwave Measurement Conf. Fall, 157-165, 2002.

6. Ferrero, A. and U. Pisani, "Two-port network analyzer calibration using an unknown `thru'," IEEE Microwave Guided Wave Letter, Vol. 2, No. 12, 505-507, 1992.
doi:10.1109/75.173410

7. Marks , R. B., "A multiline method of network analyzer calibration," IEEE Trans. on Microw. Theory and Tech., Vol. 39, No. 7, 1205-1215, 1991.
doi:10.1109/22.85388