FEM MODELING FOR PERFORMANCE EVALUATION OF AN ELECTROMAGNETIC ONCOLOGY DEEP HYPERTHERMIA APPLICATOR WHEN USING MONOPOLE, INVERTED T, AND PLATE ANTENNAS

By C. J. Trujillo-Romero, L. Leija-Salas, and A. Vera-Hernandez

Full Article PDF (1,111 KB)

Abstract:
This study focuses on the evaluation of the performance of a rectangular waveguide for deep hyperthermia when different antennas are used. Although there are several models of hyperthermia applicators, there are no studies of the advantages of employing different antennas for waveguides used in deep seated tumor treatments. Monopole antennas are the most used radiating elements inside waveguides. Here, the modeling of a monopole and two new proposed antennas, inverted T and plate, in order to find their optimal performance is presented. Parameters like output power, SWR and transmission coefficient generated for each modeled antenna were calculated by using the finite element method. The antennas with the best performance were selected in order to model an applicator-phantom system, which was used to calculate the temperature distributions generated inside the muscle phantom. The models were based on Maxwell and bioheat equations. Finally, thermal distributions were obtained and compared. The results indicate that the plate antenna generated a better focusing. The SWR obtained was 1.25, the output power was 54.71 W of 66 W applied, and the 42°C isotherm had a size of 2 cm x 2 cm.

Citation:
C. J. Trujillo-Romero, L. Leija-Salas, and A. Vera-Hernandez, "FEM Modeling for Performance Evaluation of an Electromagnetic Oncology Deep Hyperthermia Applicator When Using Monopole, Inverted T, and Plate Antennas," Progress In Electromagnetics Research, Vol. 120, 99-125, 2011.
doi:10.2528/PIER11071809
http://www.jpier.org/PIER/pier.php?paper=11071809

References:
1. Gardner, R. A., H. I. Vargas, J. B. Block, C. L. Vogel, A. J. Fenn, G. V. Kuehl, and M. Doval, "Focused microwave phased array thermotherapy for primary breast cancer," Annals of Surgical Oncology, Vol. 9, 326-332, May 2002.
doi:10.1007/BF02573866

2. Lagendijk, J. J., "Hyperthermia treatment planning," Phys. Med. Biol., Vol. 45, R61-76, May 2000.
doi:10.1088/0031-9155/45/5/201

3. Sabariego, R. V., L. Landesa, and F. Obelleiro, "Design of a microwave array hyperthermia applicator with a semicircular reflector ," Med. Biol. Eng. Comput., Vol. 37, 612-617, Sep. 1999.
doi:10.1007/BF02513356

4. Gupta, R. C. and S. P. Singh, "Elliptically bent slotted waveguide conformal focused array for hyperthermia treatment of tumors in curved region of human body," Progress In Electromagnetics Research, Vol. 62, 107-125, 2006.
doi:10.2528/PIER06012801

5. Cheung, A. Y. and A. Neyzari, "Deep local hyperthermia for cancer therapy: external electromagnetic and ultrasound techniques," Cancer. Res., Vol. 44, 4736s-4744s, Oct. 1984.

6. Kato, H. and T. Ishida, "Present and future-status of noninvasive selective deep heating using RF in hyperthermia," Med. Biol. Eng. Comput., Vol. 31, S2-S11, Jul. 1993.
doi:10.1007/BF02446643

7. Wiersma, J. and J. D. P. Van Dijk, "RF hyperthermia array modelling; validation by means of measured EM-field distributions," Int. J. Hyperthermia, Vol. 17, 63-81, Jan. 2001.
doi:10.1080/02656730150201606

8. Nilsson, P., T. Larsson, and B. Persson, "Absorbed power distributions from two tilted waveguide applicators," Int. J. Hyperthermia, Vol. 1, 29-43, Jan.-Mar. 1985.
doi:10.3109/02656738509029272

9. Chen, Z. N., K. Hirasawa, and K. Wu, "A broad-band sleeve monopole integrated into parallel-plate waveguide," IEEE Transactions on Microwave Theory and Techniques, Vol. 48, 1160-1163, Jul. 2000.
doi:10.1109/22.848502

10. Park, M. Y. and H. J. Eom, "Analysis of a coaxially fed monopole in a rectangular waveguide," IEEE Microwave and Wireless Components Letters, Vol. 15, 253-255, Apr. 2005.
doi:10.1109/LMWC.2005.845732

11. Bialkowski, M. E., "On the link between top-hat monopole antennas, disk-resonator diode mounts and coaxial-to-waveguide transitions," IEEE Transactions on Antennas and Propagation, Vol. 48, 1011-1013, Jun. 2000.
doi:10.1109/8.865244

12. Bialkowski, M. E., "Analysis of a coaxial-to-wave-guide adapter including a discended probe and a tuning post," IEEE Transactions on Microwave Theory and Techniques, Vol. 43, 344-349, Feb. 1995.
doi:10.1109/22.348094

13. Bui, V. P., X.-C. Wei, and E. P. Li, "An efficient simulation technology for characterizing the ultra-wide band signal propagation in a wireless body area network," Journal of Electromagnetic Waves and Applications, Vol. 24, No. 17-18, 2575-2588, 2010.
doi:10.1163/156939310793675691

14. Paulsen, K. D., Calculation of Power Deposition Patterns in Hyperthermia, Clinical Thermology: Thermal Modeling and Thermal Dosimetry, Vol. 2, Springer-Verlag, 1988.

15. Marcuvitz, N., The Institution of Engineering and Technology, Waveguide Handbook.

16. Hussain, A. and Q. A. Naqvi, "Fractional rectangular impedance waveguide," Progress In Electromagnetics Research, Vol. 96, 101-116, 2009.
doi:10.2528/PIER09060801

17. Gabriele, P., T. Ferrara, B. Baiotto, E. Garibaldi, P. G. Marini, G. Penduzzu, V. Giovannini, F. Bardati, and C. Guiot, "Radio hyperthermia for re-treatment of superficial tumours," Int. J. Hyperthermia, Vol. 25, 189-98, May 2009.
doi:10.1080/02656730802669593

18. Gong, Y. and G. Wang, "Superficial tumor hyperthermia with flat left-handed metamaterial lens," Progress In Electromagnetics Research, Vol. 98, 389-405, 2009.
doi:10.2528/PIER09091401

19. Iero, D., T. Isernia, A. F. Morabito, I. Catapano, and L. Crocco, "Optimal constrained field focusing for hyperthermia cancer therapy: A feasibility assessment on realistic phantoms," Progress In Electromagnetics Research, Vol. 102, 125-141, 2010.
doi:10.2528/PIER10011207

20. Carr, J. J., Practical Antenna Handbook, 4th edition, Mcgraw-Hill Professional, 2001.

21. Kumaradas, J. C. and M. D. Sherar, "An edge-element based finite element model of microwave heating in hyperthermia: application to a bolus design," Int. J. Hyperthermia, Vol. 18, 441-53, Sep.-Oct. 2002.

22. Van Rhoon, G. C., P. J. Rietveld, and J. Van Der Zee, "A 433MHz Lucite cone waveguide applicator for superficial hyperthermia," Int. J. Hyperthermia, Vol. 14, 13-27, Jan.-Feb. 1998.

23. Frickey, D. A., "Conversions between S, Z, Y , H, ABCD, and T parameters which are valid for complex source and load impedances," IEEE Transactions on Microwave Theory and Techniques, Vol. 42, 205-211, 1994.
doi:10.1109/22.275248

24. Liang, C.-H., Y. Shi, and T. Su, "S parameter theory of lossless block network," Progress In Electromagnetics Research, Vol. 104, 253-266, 2010.
doi:10.2528/PIER10022611

25. Pennes, H. H., "Analysis of skin, muscle and brachial arterial blood temperatures in the resting normal human forearm," Am. J. Med. Sci., Vol. 215, 354, Mar. 1948.

26. Mohsin, S. A., N. M. Sheikh, and W. Abbas, "MRI induced heating of artificial bone implants," Journal of Electromagnetic Waves and Applications, Vol. 23, No. 5-6, 799-808, 2009.
doi:10.1163/156939309788019796

27. Ebrahimi-Ganjeh, M. A. and A. R. Attari, "Study of water bolus effect on SAR penetration depth and effective field size for local hyperthermia," Progress In Electromagnetics Research B, Vol. 4, 273-283, 2008.
doi:10.2528/PIERB08011403

28. Gabriel, C., "Compilation of the dielectric properties of body tissues at RF and microwave frequencies," , Report N.AL/OE-TR-1996-0037, Occupational and environmental heal-th directorate, Radiofrequency Radiation Division, Brooks Air Force Base, Texas (USA), June 1996.